Mayo Test ID NAU Sodium, 24 Hour, Urine
Necessary Information
24-Hour volume (in milliliters) is required.
Specimen Required
Supplies: Sarstedt 5 mL Aliquot Tube (T914)
Collection Container/Tube: 24-hour graduated urine container with no metal cap or glued insert
Submission Container/Tube: Plastic tube or a clean, plastic aliquot container with no metal cap or glued insert
Specimen Volume: 5 mL
Collection Instructions:
1. Collect urine for 24 hours.
2. Refrigerate specimen within 4 hours of completion of 24-hour collection.
Additional Information: See Urine Preservatives-Collection and Transportation for 24-Hour Urine Specimens for multiple collections.
Useful For
Assessing acid-base balance, water balance, water intoxication, and dehydration
Special Instructions
Method Name
Indirect Potentiometry
Reporting Name
Sodium, 24 HR, USpecimen Type
UrineSpecimen Minimum Volume
1 mL
Specimen Stability Information
| Specimen Type | Temperature | Time |
|---|---|---|
| Urine | Refrigerated (preferred) | 14 days |
| Frozen | 30 days | |
| Ambient | 7 days |
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Clinical Information
Sodium (Na+) is the primary extracellular cation. Sodium is responsible for almost one half the osmolality of the plasma and, therefore, plays a central role in maintaining the normal distribution of water and the osmotic pressure in the extracellular fluid compartment. The amount of Na+ in the body is a reflection of the balance between Na+ intake and output. The normal daily diet contains 8 to 15 grams of sodium chloride (NaCl) which is nearly completely absorbed from the gastrointestinal tract. The body requires only 1 to 2 mmol/day, and the excess is excreted by the kidneys, which are the ultimate regulators of the amount of Na+ (and thus water) in the body. Sodium is freely filtered by the glomeruli. Approximately 70% to 80% of the filtered Na+ is actively reabsorbed in the proximal tubules, with chloride and water passively following in an iso-osmotic and electrically neutral manner. Another 20% to 25% is reabsorbed in the loop of Henle along with chloride and more water. In the distal tubules, interaction of the adrenocortical hormone aldosterone with the coupled sodium-potassium and sodium-hydrogen exchange systems directly results in the reabsorption of Na+ and indirectly of chloride from the remaining 5% to 10% of the filtered load. It is the regulation of this latter fraction of filtered Na+ that determines the amount of Na+ excreted in the urine.
Reference Values
≥18 years: 22-328 mmol/24 hours
Reference values have not been established for patients who are less than 18 years of age.
Interpretation
Urinary sodium (Na+) excretion varies with dietary intake, and there is a large diurnal variation with the rate of Na+ excretion during the night, being only 20% of the peak rate during the day.
Sodium may be lost in the kidneys as a result of diuretic therapy, salt-losing nephropathies, or adrenal insufficiency, with the urinary Na+ concentration usually more than 20 mEq/L. In these hypovolemic states, urine Na+ values <10 mEq/L indicate extrarenal Na+ loss. In hypervolemic states, a low urine Na+ (<10 mEq/L) may indicate nephrotic syndrome in addition to non-kidney causes.
Cautions
No significant cautionary statements.
Clinical Reference
Delaney MP, Lamb EJ: Kidney disease. In: Rifai NF, Horvath AR, Wittwer CT, eds. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. 6th ed. Elsevier; 2018:1308-1309
Method Description
An ion-selective electrode (ISE) makes use of the unique properties of ion-selective membrane to develop an electrical potential (electromotive force: EMF) for the measurements of ions in solution. Selective membrane is in contact with both the test solution and an internal filling solution. Due to the selectivity of the membrane, only the ions to be measured contribute to the EMF. The membrane EMF is determined by the difference in concentration of the test ion in the test solution and the internal filling solution. The EMF develops according to Nernst equation of a specific ion in solution.(Package insert: ISE indirect Na-K-Cl for Gen 2. Roche Diagnostics; 05/2024)
Specimen Retention Time
1 weekPerforming Laboratory
Mayo Clinic Laboratories in Rochester
Test Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
84300
LOINC Code Information
| Test ID | Test Order Name | Order LOINC Value |
|---|---|---|
| NAU | Sodium, 24 HR, U | 2956-1 |
| Result ID | Test Result Name | Result LOINC Value |
|---|---|---|
| NA_24 | Sodium, 24 HR, U | 2956-1 |
| TM11 | Collection Duration | 13362-9 |
| VL9 | Urine Volume | 3167-4 |
Day(s) Performed
Monday through Sunday