Mayo Test ID SGSU Synthetic Glucocorticoid Screen, Random, Urine
Ordering Guidance
Specimen Required
Supplies: Urine tubes, 10 mL (T068)
Container/Tube: Plastic, 10-mL urine tube
Specimen Volume: 5 mL
Collection Instructions:
1. Collect a random urine specimen.
2. No preservative.
Useful For
Confirming the presence of the listed synthetic glucocorticoids
Confirming the cause of secondary adrenal insufficiency
Method Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Reporting Name
Synthetic Glucocorticoid Screen, USpecimen Type
UrineSpecimen Minimum Volume
0.6 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Urine | Frozen | 14 days |
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Clinical Information
Synthetic glucocorticoids are widely used and have important clinical utility both as anti-inflammatory and immunosuppressive agents. The medical use of these agents, as well as their surreptitious use, can sometimes lead to a confusing clinical presentation. Patients exposed to these steroids may present with clinical features of Cushing syndrome but with suppressed cortisol levels and evidence of hypothalamus-pituitary-adrenal axis suppression.
Reference Values
Negative
Cutoff concentrations
Betamethasone: 0.10 mcg/dL
Budesonide: 0.20 mcg/dL
Dexamethasone: 0.10 mcg/dL
Fludrocortisone: 0.10 mcg/dL
Megestrol acetate: 0.10 mcg/dL
Methylprednisolone: 0.10 mcg/dL
Prednisolone: 0.10 mcg/dL
Prednisone: 0.10 mcg/dL
Triamcinolone acetonide: 0.10 mcg/dL
Values for normal patients not taking these synthetic glucocorticoids should be less than the cutoff concentration (detection limit).
Interpretation
This test screens for and quantitates, if present, the following synthetic glucocorticoids: betamethasone, budesonide, dexamethasone, fludrocortisone, megestrol acetate, methylprednisolone, prednisolone, prednisone, and triamcinolone acetonide.
The presence of synthetic glucocorticoids in urine indicates current or recent use of these compounds. Since several of these compounds exceed the potency of endogenous cortisol by 1 or more orders of magnitude, even trace levels may be associated with cushingoid features.
Cautions
This method cannot detect all synthetic steroids available either as pharmaceutical compounds or chemicals present in food. The assay confirms only the listed synthetic glucocorticoids. For more information see Interpretation.
Lack of detection does not preclude use of synthetic glucocorticoid because adrenal suppression may persist for some time after the exogenous steroid is discontinued.
Clinical Reference
1. Cave A, Arlett P, Lee E. Inhaled and nasal corticosteroids: factors affecting the risks of systemic adverse effects. Pharmacol Ther. 1999;83(3):153-179
2. Bijlsma JWJ, Van Everdingen AA, Huisman M, De Nijs RNJTL, Jacobs JWG. Glucocorticoids in rheumatoid arthritis: effects on erosions and bone. Ann NY Acad Sci. 2002;966:82-90
3. Sandborn WJ. Steroid-dependent Crohn's disease. Can J Gastroenterol. 2000;14 Suppl C:17C-22C
4. Benvenuti S, Brandi ML. Corticosteroid-induced osteoporosis: pathogenesis and prevention. Clin Exp Rheumatol. 2000;18(4 Suppl 20):S64-S66
5. Loke TK, Sousa AR, Corrigan CJ, Lee TH. Glucocorticoid-resistant asthma. Curr Allergy Asthma Rep. 2002;2(2):144-150
6. Fardet L, Petersen I, Nazareth I. Monitoring of patients on long-term glucocorticoid therapy: a population-based cohort study. Medicine (Baltimore). 2015;94(15):e647. doi:10.1097/MD.0000000000000647
7. Cronin JJ, McCoy S, Kennedy U, et al. A randomized trial of single-dose oral dexamethasone versus multidose prednisolone for acute exacerbations of asthma in children who attend the emergency department. Ann Emerg Med. 2016;67(5):593-601.e3. doi:10.1016/j.annemergmed.2015.08.001
Method Description
The synthetic glucocorticoids are extracted from 0.5 mL of urine using an acetonitrile protein precipitation followed by methylene chloride liquid extraction of the solvent. Cortisol-9, 11, 12, 12-d, and triamcinolone-d1 acetonide-d6 are added to each sample before the liquid extraction and serve as the internal standards. Then, 17 mcL of the reconstituted sample extract is injected into a high-performance liquid chromatography system and analyzed by tandem mass spectrometry. The mass spectrometer has an electrospray interface and is operated in the multiple reaction monitoring positive mode. The calibration utilizes a 4-point standard curve over a concentration range of 0 to 25 mcg/dL.(McWhinney BC, Ward G, Hickman PE. Improved HPLC method for simultaneous analysis of cortisol, 11-deoxycortisol, prednisolone, methylprednisolone, and dexamethasone in serum and urine. Clin Chem. 1996;42:979-981; Savu S, Silvestro L, Haag A, Sorgel F. A confirmatory HPLC-MS/MS method for ten synthetic corticosteroids in bovine urines. J Mass Spectrom. 1996;31[12]:1351-1363; Djedovic NK, Rainbow SJ. Detection of synthetic glucocorticoids by liquid chromatography-tandem mass spectrometry in patients being investigated for Cushing's syndrome. Ann Clin Biochem. 2011;48[Pt 6]:542-549. doi:10.1258/acb.2011.010250)
Day(s) Performed
Wednesday
Report Available
5 to 11 daysSpecimen Retention Time
3 monthsPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
80299
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
SGSU | Synthetic Glucocorticoid Screen, U | 46959-3 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
23562 | Betamethasone | 46946-0 |
23563 | Budesonide | 46947-8 |
23564 | Dexamethasone | 46948-6 |
23565 | Fludrocortisone | 46949-4 |
23569 | Megestrol Acetate | 46953-6 |
23570 | Methylprednisolone | 46954-4 |
23571 | Prednisolone | 46955-1 |
23572 | Prednisone | 46956-9 |
23574 | Triamcinolone Acetonide | 46958-5 |